VISCOMETER Tube, Canon-Fenske method

MAS 5131- 5132 – 5133 – 5133- 5134- 5135- 5136- 5137- 5138- 5139-5140-5141-5142-5142-5143-5144

ASTM D445, ASTM D2170, ISO 3104

The Cannon-Fenske Routine Viscometer measures kinematic viscosity of transparent Newtonian liquids, particularly petroleum products or lubricants, according to ASTM D445 and ISO 3104. Specifications conform to ASTM D446 and ISO 3105. Minimum sample volume is 7 mL. Minimum bath depth is 230 mm (8 in). Special constant viscometers are available at additional cost. Available as a calibrated viscometer (which includes a certificate of calibration and instruction sheet) or an uncalibrated viscometer. Holders are sold separately.

CANNON-Fenske RoutineViscometers

CATALOG NUMBERSIZEAPPROXIMATE CONSTANT
(MM²/S²)
VISCOSITY RANGE (MM²/S)†
UNCALIBRATEDCALIBRATEDLOWHIGH
MAS-5131AMAS-5131B250.0020.52.0
MAS-5132AMAS-5132B500.0040.84.0
MAS-5133AMAS-5133B750.0081.68.0
MAS- 5134AMAS-5134B1000.0153.015
MAS-5135AMAS-5135B1500.0357.035
MAS-5136AMAS-5136B2000.120100
MAS-5137AMAS-5137B3000.2550250
MAS-5138AMAS-5138B3500.5100500
MAS-5139AMAS-5139B4001.22401200
MAS-5140AMAS-5140B4502.55002500
MAS-5141AMAS-5141B5008.016008000
MAS-5142AMAS-5142B60020400020000
MAS-5143AMAS-5143B65045900045000
MAS-5144AMAS-5144B70010020000100000

Viscosity ranges in mm²/s represent flow times of 250 to 1000 seconds for size 25 and 200 to 1000 seconds for all other sizes