Pensky-Martens Closed Cup Flash Point Tester For Petroleum – MAS 5464-LYD-261

Closed cup flash point tester is used primarily to test flammable and combustible materials for shipping and safety regulations. Consists of test flame applicator, brass test cup, thermometer support, heating plate and electric heater. Applicator is precisely aligned per specifications and pivots for test flame application at specified temperature intervals.

SPECIFICATIONS

ModelLYD-261
Power supplyAC (220±10%)V, 50Hz
Total power consumption≤650W
Heating deviceThe furnace body is made of   silicon carbide
Heating power is adjustable   from (0-600) w
Heating modeAdjust by manual
Stir modeMechanical drive stirring
Stirring rateProcedure A: (90~120)RPMProcedure B: (250±10)RPM
Oil cupInner diameter50.7mm~50.8mm
Depth55.7mm~56.0mm
The scribed line depth of   capacity of testing oil33.9mm~34.3mm
Capacity of testing oilabout 70ml
Igniting sourcegas(or   other civilian fuels,the same below)
Mercury-in-glass   thermometerScale -5℃~110℃,   division 0.5℃Scale 20℃~150℃,   division 1℃Scale 90℃~370℃,division   2℃
Ambient temperature(15~35)℃
Relative humidity≤ 85%
Dimension370mm×320mm×300mm
Net weight7kg

DESCRIPTION

This instrument is designed and manufactured as per standards GB/T 261-2008 Determination of Flash Point – Pensky-Martens Closed Cup Method and ASTM D93 Standard Test Method for Flash Point by Pensky-Martens Closed Cup Tester. It is used to make determination of the closed cup flash point of the petroleum products.

FEATURES

1. This instrument is designed as per GB/T261 and ASTM D93.

2. The heating power is continuously adjustable. The power is shown visually by voltmeter. The temperature control mode is advanced and reasonable.

3. The structure is designed small and exquisite. The stainless-steel table board is beautiful and easy to clean.

4. Operation is easy. Test results are precise. It has been chosen as the instrument to test the closed cup flash point by many metering and detecting institutes.